EasyManua.ls Logo

Rohde & Schwarz RTE Series - Noise Setup

Rohde & Schwarz RTE Series
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1939User Manual 1326.1032.02 ─ 20
Parameters:
<Inversion> ON | OFF
*RST: OFF
WGENerator<m>:VOLTage:LOW <Low>
Sets the low signal level of the output waveform.
Suffix:
<m>
.
1..2
Parameters:
<Low> Range: -6 to 5.99
Increment: 0.1
*RST: -0.5
Default unit: V
WGENerator<m>:VOLTage:OFFSet <Offset>
Sets a voltage offset.
Suffix:
<m>
.
1..2
Parameters:
<Offset> Range: -5.995 to 5.995
Increment: 0.1
*RST: 0
Default unit: V
18.19.4 Noise setup
WGENerator<m>:MODulation:NDCLevel........................................................................1939
WGENerator<m>:MODulation:NLABsolute?................................................................... 1940
WGENerator<m>:MODulation:NLPCent......................................................................... 1940
WGENerator<m>:MODulation:NOISe.............................................................................1940
WGENerator<m>:MODulation:NDCLevel <LevelDC>
Sets the DC noise level.
Suffix:
<m>
.
1..2
Parameters:
<LevelDC> Range: 0 to 12
Increment: 0.1
*RST: 0.01
Default unit: V
Waveform generator (option R&S RTE-B6)

Table of Contents

Other manuals for Rohde & Schwarz RTE Series

Related product manuals