Waveform generator (Option R&S RTE-B6)
R&S
®
RTE
948User Manual 1326.1032.02 ─ 20
Remote command:
WGENerator<m>:COUPling:OUTPut on page 1942
Couple all frequency parameters
Couples all frequency parameters of "Gen1" to "Gen2" or "Gen2" to "Gen1". The val-
ues are taken from the currently selected generator.
For "Operation Mode" > "Function Gen.": "Frequency" and "Period"
For "Operation Mode" > "Modulation": "Carrier frequency", "Carrier period", "Modu-
lation freq", "Freq deviation", "Frequency 1", "Frequency 2" and "Hop rate"
For "Operation Mode" > "Sweep": "Start frequency", "Stop frequency" and "Sweep
time".
For "Operation Mode" > "Arbitrary": "Arb wfm source" and "Signal source".
You can still change the other settings of the generators independently.
Remote command:
WGENerator<m>:COUPling:AMPLitude on page 1942
Phase shift
Sets the phase shift between the waveform of "Gen1" and "Gen2" when the frequency
parameters of the two waveforms are coupled.
Remote command:
WGENerator<m>:COUPling:PHASeshift on page 1942
Couple all amplitude parameters except Inversion and Load
Couples all amplitude parameters of "Gen1" to "Gen2" or "Gen2" to "Gen1" including
the load. The values are taken from the currently selected generator.
For all "Operation Mode": the output settings except of "Load" and "Inversion",
"Amplitude", "High", "Offset" and "Low".
For "Operation Mode" > "Function Gen.": "DC Level"
For "Operation Mode" > "Modulation": "Modulation depth"
Remote command:
WGENerator<m>:COUPling[:FREQuency] on page 1942
Sync
Selects, which signals generated from the waveform generator are synchronized.
Selecting one of the sync options indicates that the first samples of those signals are
generated at the same time, irrespective of if the generators are on or off. Selecting
one of the coupling options automatically syncs the signals generated by the two wave-
form generators.
Remote command:
GENerator:SYNC[:COMBination] on page 1943