Remote control commands
R&S
®
RTE
1989User Manual 1326.1032.02 ─ 20
STATus:QUEStionable:ZVCoverload:PTRansition <Value>
STATus:QUEStionable:MASK:PTRansition <Value>
Sets the positive transition filter. If a bit is set, a 0 to 1 transition in the corresponding
bit of the condition register causes a 1 to be written in the corresponding bit of the
event register.
Parameters:
<Value> Bit mask in decimal representation
Example:
STATus:QUEStionable:MASK:PTRansition 24
Set bits no. 3 and 4 of the STATus:QUEStionable:MASK:PTRan-
sition register part: 24 = 8 + 16 = 2
3
+ 2
4
MTESt:SBITnumber? <MaskTestName>
Returns the number of the status bit that belongs to the indicated mask test in the
STATus:QUEStionable:MASK register. The status bits are assigned in the order of
mask creation.
Query parameters:
<MaskTestName> String with the name of the mask test
Return values:
<StatusBitNumber> Bit number, see "STATus:QUEStionable:MASK register"
on page 1051.
Range: 0 to 7
Increment: 1
*RST: 0
Example:
:MTESt:SBITnumber? 'MT1'
<--0
:MTESt:SBITnumber? 'MT2'
<--1
:MTESt:SBITnumber? 'MT3'
<--2
Usage: Query only
Asynchronous command
18.23 Remote trace
The commands in this section configure tracing of the remote control interface and of
events. They also configure the display of the SCPI remote trace.
Before you start tracing, configure all settings as desired. Modifying settings while trac-
ing is active may result in loss of already traced data. Useful exception: Selecting a
new target file while tracing is allowed. For start mode EXPLicit a restart of the
instrument resets the settings to the documented default values.