Remote control commands
R&S
®
RTE
1558User Manual 1326.1032.02 ─ 20
Query parameters:
<SearchName>
Return values:
<ByteValue> Range: 0 to 255
Increment: 1
*RST: 0
Usage: Query only
18.17.8 Flexray (option R&S RTE-K4)
Configuration.......................................................................................................1558
Trigger.................................................................................................................1562
Decode results.................................................................................................... 1571
Search settings................................................................................................... 1577
Search results..................................................................................................... 1586
18.17.8.1 Configuration
BUS<m>:FLXRay:SRCType..........................................................................................1558
BUS<m>:FLXRay:SOURce<n>..................................................................................... 1559
BUS<m>:FLXRay:THReshold<n>..................................................................................1559
BUS<m>:FLXRay:THENable.........................................................................................1560
BUS<m>:FLXRay:THData............................................................................................ 1560
BUS<m>:FLXRay:PRSingle.......................................................................................... 1560
BUS<m>:FLXRay:PRDiff.............................................................................................. 1561
BUS<m>:FLXRay:PRLogic........................................................................................... 1561
BUS<m>:FLXRay:POLarity...........................................................................................1561
BUS<m>:FLXRay:BITRate............................................................................................1562
BUS<m>:FLXRay:CHTYpe........................................................................................... 1562
BUS<m>:FLXRay:SEHB...............................................................................................1562
BUS<m>:FLXRay:SRCType <SourceType>
Sets the type of measurement.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<SourceType> SINGle | DIFFerential | LOGic
SINGle
Used for measurements with single-ended probes or single-
ended voltage measurements with differential probes on the
FlexRay bus. Two thresholds have to be defined as absolute
voltage levels, see BUS<m>:FLXRay:THReshold<n>
on page 1559.