EasyManua.ls Logo

Rohde & Schwarz RTE Series - Triggering on Ethernet Signals

Rohde & Schwarz RTE Series
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Protocol analysis
R&S
®
RTE
702User Manual 1326.1032.02 ─ 20
"Length/Typ
(Max)"
Defines the length/type pattern that is required to specify a range with
conditions "In range" and "Out of range".
Remote command:
TRIGger<m>:ETHernet:FRAMe:TCONdition on page 1656
TRIGger<m>:ETHernet:FRAMe:TMAX on page 1656
TRIGger<m>:ETHernet:FRAMe:TMIN on page 1656
Frame check
Defines the frame check sequence value.
"Condition"
Sets the comparison condition to a specific value or a range.
"Frame check
(Min)"
Specifies the value or sets the start value of a range.
"Frame check
(Max)"
Sets the maximum value of a range for "Condition" = "In range"/"Out
of range".
Remote command:
TRIGger<m>:ETHernet:FRAMe:CCONdition on page 1653
TRIGger<m>:ETHernet:FRAMe:CMAX on page 1654
TRIGger<m>:ETHernet:FRAMe:CMIN on page 1654
Error
Triggers on enabled errors.
"Preamble
error"
Triggers on a frame with invalid preamble.
"Length error"
Triggers on an incorrect length of the sequence - when additional or
missing bits are detected and the sequence of bits is not as expected.
"CRC error"
Triggers on a frame that has a mismatch of the cyclic redundancy
check (CRC) value between the transmitting and receiving device.
Remote command:
TRIGger<m>:ETHernet:ERRor:CRC on page 1652
TRIGger<m>:ETHernet:ERRor:LENGth on page 1653
TRIGger<m>:ETHernet:ERRor:PREamble on page 1653
13.11.3.2 Triggering on Ethernet signals
Prerequisite: A bus is configured for the Ethernet signal to be analyzed.
For the basic trigger settings, proceed in the following way:
1. Press the [PROTOCOL] key and select the "Trigger" tab.
2. Press the "Select" button. The "Source" button is shown.
3. Tap the "Source" button and select the "Serial bus" trigger source.
4. Select the serial bus that is set to Ethernet.
5. Tap "Type" and select the trigger type to be used for Ethernet protocol analysis.
Available trigger types are "Frame Start", "Pattern", "Adv frame" and "Adv error".
Ethernet 10BASE-T and 100BASE-TX (option R&S
RTE-K8)

Table of Contents

Other manuals for Rohde & Schwarz RTE Series

Related product manuals