Remote control commands
R&S
®
RTE
1108User Manual 1326.1032.02 ─ 20
Usage: Event
Asynchronous command
18.8.2 Time base
TIMebase:SCALe.........................................................................................................1108
TIMebase:RANGe........................................................................................................1108
TIMebase:DIVisions?....................................................................................................1109
TIMebase:HORizontal:POSition.....................................................................................1109
TIMebase:REFerence...................................................................................................1109
TRIGger<m>:OFFSet:LIMited........................................................................................1109
AUToscale................................................................................................................... 1110
ACQuire:POINts:AUTO................................................................................................. 1110
ACQuire:POINts:AADJust..............................................................................................1110
ACQuire:POINts:MAXimum........................................................................................... 1111
ACQuire:POINts:ARATe?...............................................................................................1111
ACQuire:SRATe............................................................................................................1111
ACQuire:SRReal.......................................................................................................... 1112
ACQuire:RESolution..................................................................................................... 1112
ACQuire:POINts[:VALue]...............................................................................................1112
TIMebase:ROLL:ENABle...............................................................................................1112
TIMebase:ROLL:STATe?............................................................................................... 1113
TIMebase:ROLL:MTIMe................................................................................................1113
TIMebase:SCALe <TimeScale>
Sets the horizontal scale - the time per division on the x-axis - for all channel and math
waveforms.
The setting accuracy depends on the current resolution (sample rate).
No interpolation:
The resolution is an integer multiple of the ADC sample rate.
With interpolation:
Any value for the horizontal scale can be set.
Parameters:
<TimeScale> Range: 25E-12 to 10000 (RTO, RTP) | 5000 (RTE)
Increment: 1E-12
*RST: 10E-9
Default unit: s/div
Usage: Asynchronous command
TIMebase:RANGe <AcquisitionTime>
Defines the time of one acquisition, that is the time across the 10 divisions of the dia-
gram: TimeScale*10.