Probe skew offset
Measures the skew of all connected active probes and includes it in the total skew off-
set.
"Use probe group delay ..."
If enabled, the skew of all connected active probes is measured, dis-
played, and used for deskewing. The setting affects all active chan-
nels.
"Active probe"
Shows the type of the probe that is connected to the selected chan-
nel.
"Probe group
delay"
Shows the result of the probe skew measurement on the selected
channel.
Remote command:
PROBe<m>:SKEState on page 1162
Total skew offset
Shows the effective skew offset, the sum of the measured "Probe group delay" and the
"Skew offset". If "Use skew offset" is disabled, the skew offset is ignored.
5.9.2 Reference clock
Access: "Horizontal" menu > "Skew" > "Reference clock" tab.
Input and output reference signals are connected to the [REF IN/OUT] connector on
the rear panel of R&S RTE. You can select an internal or external reference clock on
the "Reference clock" tab.
Reference
Sets the reference clock that is to be used.
"Internal"
Uses the internal reference signal.
"External"
Uses an external reference signal.
Remote command:
SENSe[:ROSCillator]:SOURce on page 1163
External reference
Sets the frequency of an external reference input signal: 10 MHz.
Remote command:
SENSe[:ROSCillator]:EXTernal:FREQuency on page 1163