Remote control commands
R&S
®
RTE
1158User Manual 1326.1032.02 ─ 20
Usage: Asynchronous command
DIFFerential<m>:PSIGnal[:SELect] <Signal>
DIFFerential<m>:NSIGnal[:SELect] <Signal>
Select the analog input channel of the positive and negative signal.
For differential signal 1, analog Ch1 and Ch3 are used. Differential signal 2 uses Ch2
and Ch4.
Suffix:
<m>
.
1..2
Differential signal
Parameters:
<Signal> C1W1 | C2W1 | C3W1 | C4W1
*RST: C1W1
Usage: Asynchronous command
DIFFerential<m>:AOUTput <Output>
DIFFerential<m>:BOUTput <Output>
Select the waveform that is the result of differential processing. 2 output waveforms per
digital signal can be defined.
Suffix:
<m>
.
1..2
Differential signal
Parameters:
<Output> P | N | DIFF | COM
P | N
Positive or negative waveform.
DIFF
Differential waveform.
COM
Common mode waveform.
*RST: COM
Usage: Asynchronous command
DIFFerential<m>:COMMon:SCALe <Scale>
DIFFerential<m>:DIFFerential:SCALe <Scale>
Set the vertical scale of differential and common mode waveforms, respectively.
Vertical settings of P and N output are the vertical settings of the input channels.
Suffix:
<m>
.
1..2
Differential signal
Acquisition and setup