Remote control commands
R&S
®
RTE
1114User Manual 1326.1032.02 ─ 20
<n> 1..3
[:WAVeform<n>] is irrelevant, omit it.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CHAN2 ON
Activates analog channel 2.
Usage: Asynchronous command
CHANnel<m>[:WAVeform<n>]:TYPE <DecimationMode>
Selects the method to reduce the data stream of the ADC to a stream of waveform
points with lower sample rate.
See also:"Decimation" on page 181.
Suffix:
<m>
.
1..4
Selects the input channel.
<n> 1..3
[:WAVeform<n>] is irrelevant, omit it.
Parameters:
<DecimationMode> SAMPle | PDETect | HRESolution | RMS
SAMPle
One of n samples in a sample interval of the ADC is recorded as
waveform point.
PDETect
Peak Detect: the minimum and the maximum of n samples in a
sample interval are recorded as waveform points.
HRESolution
High resolution: The average of n sample points is recorded as
waveform point.
RMS
The waveform point is the root mean square of n sample values.
*RST:
SAMPle
Usage: Asynchronous command
CHANnel<m>[:WAVeform<n>]:ARIThmetics <Arithmetics>
Selects the method to build the resulting waveform from several consecutive acquisi-
tions of the signal. To define the number of acquisitions, use ACQuire:COUNt.