Remote control commands
R&S
®
RTE
1147User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1..2
<n> 1..4
Setting parameters:
<Settings> ADJusted | ORIGinal
ADJusted
The operation range that is automatically adjusted by the probe
during overload is kept.
In internal shunt mode, during an overload at a current channel,
the probe switches automatically to the next higher range. If the
10A range is also exceeded, the amperemeter switches to exter-
nal shunt mode to protect the probe against permanent damage.
ORIGinal
The original operation range is restored as before the overload
and the adjustment of the operation range.
Usage:
Setting only
Asynchronous command
ZVC:Z<m>:I<n>:OVERload:VALue?
Queries if an overload of the current channel was detected.
Suffix:
<m>
.
1..2
<n> 1..4
Return values:
<Overload> ON | OFF
*RST: OFF
Usage: Query only
Asynchronous command
ZVC:Z<m>:I<n>:POSition <VertPosi>
Sets the vertical position of the indicated current channel as a graphical value.
Suffix:
<m>
.
1..2
<n> 1..4
Parameters:
<VertPosi> Range: -5 to 5
Increment: 0.02
*RST: 0
Default unit: div
Usage: Asynchronous command