Remote control commands
R&S
®
RTE
1205User Manual 1326.1032.02 ─ 20
TRIGger<m>:SEQuence:DELay <Delay>
Sets the time the instrument waits after an A-event until it recognizes B-events.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<Delay> Range: 0 to 50
Increment: 1E-12
*RST: 0
Default unit: s
Usage: Asynchronous command
TRIGger<m>:SEQuence:COUNt
<Events>
Sets the number of B-events to be fulfilled after an A-event. The last B-event causes
the trigger.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<Events> Range: 1 to 2147483647
Increment: 1
*RST: 1
Usage: Asynchronous command
TRIGger<m>:SEQuence:RESet:EVENt <EnabRstEvt>
If set to ON, the trigger sequence is restarted by the R-event if the specified number of
B-event does not occur.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<EnabRstEvt> ON | OFF
*RST: OFF
Usage: Asynchronous command
TRIGger<m>:SEQuence:RESet:TIMeout[:ENABle] <State>
If set to ON, the instrument waits for the time defined using TRIGger<m>:SEQuence:
RESet:TIMeout:TIME for the specified number of B-events. If no trigger occurs dur-
ing that time, the sequence is restarted with the A-event.