Remote control commands
R&S
®
RTE
1229User Manual 1326.1032.02 ─ 20
MINHold
Determines the minimum result for each input value from the
math data of the current acquisition and a number of acquisi-
tions before. To define the number of acquisitions, use
ACQuire:COUNt.
*RST: OFF
Usage: Asynchronous command
CALCulate:MATH<m>:VERTical:OFFSet <VerticalOffset>
Sets a voltage offset to adjust the vertical position of the math function on the screen.
Suffix:
<m>
.
1..8
Math waveform
Parameters:
<VerticalOffset> Negative values move the waveform au, positive values move it
down.
Range: -100E+12 to 100E+12
Increment: 0.01
*RST: 0
Default unit: div
Usage: Asynchronous command
CALCulate:MATH<m>:VERTical:RANGe <VerticalRange>
Defines the range of FFT values to be displayed.
Suffix:
<m>
.
1..8
Math waveform
Parameters:
<VerticalRange> Range: 0 to 1E+15
Increment: 0.01
*RST: 0
Default unit: div
Usage: Asynchronous command
CALCulate:MATH<m>:VERTical:SCALe <VerticalScale>
Defines the scale of the y-axis in the math function diagram. The value is defined as "V
per division", e.g. 50V/div. In this case, the horizontal grid lines are displayed in inter-
vals of 50 V.
Suffix:
<m>
.
1..8
Math waveform
Waveform analysis