Remote control commands
R&S
®
RTE
1246User Manual 1326.1032.02 ─ 20
DISPlay:RESultboxes:CUPosition <Position>
Defines the position of the cursor measurement results on the screen.
Parameters:
<Position> PREV | FLOA | DOCK
PREV
Preview: result icon on the signal bar.
FLOA
Floating result box in front of the diagrams.
DOCK
Docked: fixed tab below the diagrams.
*RST:  DOCK
Usage:  Asynchronous command
CURSor<m>:XDELta[:VALue]?
Queries the delta value (distance) of two vertical cursor lines.
Suffix: 
<m>
.
1 | 2
Selects the cursor set.
Return values: 
<Delta> Range:  -100E+24  to  100E+24
*RST:  0
Default unit: s
Usage:  Query only
Asynchronous command
CURSor<m>:XDELta:INVerse?
Queries the inverse value of the delta value (distance) of the two vertical cursor lines.
Suffix: 
<m>
.
1 | 2
Selects the cursor set.
Return values: 
<DeltaInverse> Range:  -100E+24  to  100E+24
*RST:  0
Default unit: Hz
Usage:  Query only
Asynchronous command
CURSor<m>:YDELta[:VALue]?
Queries the delta value (distance) of the two horizontal cursor lines.
Cursor measurements