1255User Manual 1326.1032.02 ─ 20
Example:
:MEASurement:ARNames 1
:MEASurement:STATistics:ENABle 1
:MEASurement:ARES?
High:
1.225296442688e-001,1.225296442688e-001,
1.225296442688e-001,1.225296442688e-001,
1.225296442688e-001,1.673196471220e-014,176656,
176656
Results: current, peak+, peak-, average, RMS,
standard deviation, event count, waveform count
Usage: Query only
Asynchronous command
MEASurement<m>:ARNames
<Identifier>
Enables a prefix that indicates the measurement in the result string of the
MEASurement<m>:ARES? command.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<Identifier> ON | OFF
*RST: OFF
Example:
:MEASurement:ARNames 1
:MEASurement:ARES?
Amplitude: 2.371541501976e-001
Usage: Asynchronous command
MEASurement<m>:RESult[:ACTual]? [<MeasType>]
MEASurement<m>:RESult:AVG? [<MeasType>]
MEASurement<m>:RESult:EVTCount? [<MeasType>]
MEASurement<m>:RESult:NPEak? [<MeasType>]
MEASurement<m>:RESult:PPEak? [<MeasType>]
MEASurement<m>:RESult:RELiability? [<MeasType>]
MEASurement<m>:RESult:RMS? [<MeasType>]
MEASurement<m>:RESult:WFMCount? [<MeasType>]
MEASurement<m>:RESult:STDDev? [<MeasType>]
Return the statistic results of the selected measurement.
[:ACTual]: current measurement result
AVG: average of the long-term measurement results
EVTCount: number of measurement results in the long-term measurement
NPEak: negative peak value of the long-term measurement results
PPEak: positive peak value of the long-term measurement results