Remote control commands
R&S
®
RTE
1273User Manual 1326.1032.02 ─ 20
MEASurement<m>:SPECtrum:RESult<n>:COUNt <MaxNoOfResults>
Sets the maximum number of measurement results that are listed in the result table.
Available for peak list and harmonic search measurements (MEASurement:MAIN is
set to PLISt or HAR).
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
<n> 12 | 13
12: Peak list measurements
13: Harmonic search measurements
Parameters:
<MaxNoOfResults> Range: 1 to 1000
Increment: 1
*RST: 10
Usage: Asynchronous command
MEASurement<m>:SPECtrum:RESult<n>:MODE <ResultMode>
Sets the way how the measurement results are displayed. Available only if
MEASurement:MAIN is set to PLISt or HAR.
Suffix:
<m>
.
1..10
<n> 12 | 13
12: Peak list measurements
13: Harmonic search measurements
Parameters:
<ResultMode> ABS | REL
ABS
The harmonics/peaks are shown in absolute value, dBm.
REL
The level of the carrier is shown in absolute value dBm. The val-
ues the harmonics/peaks are shown relatively to the carrier in
dBc.
*RST: ABS
Example:
:MEASurement1:SPECtrum:RESult13:MODE ABS
The harmonic measurements are shown in absolute value, dBm.
Usage: Asynchronous command
MEASurement<m>:RESult:MAXCount <Number>
MEASurement<m>:RESult:MAXCount? <MeasType>
Defines the maximum number of peaks that are labeled in the diagram.