EasyManua.ls Logo

Rohde & Schwarz RTE Series - Page 1345

Rohde & Schwarz RTE Series
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1345User Manual 1326.1032.02 ─ 20
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:ONViolation:REPort <MaskTestName>,<Report>
MTESt:ONViolation:REPort? <MaskTestName>
Creates and saves a report of the current settings and results.
Parameters:
<Report> NOACtion | SUCCess | VIOLation
See Chapter 18.4.5, "Event parameter", on page 1075
*RST: NOACtion
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:ONViolation:TRIGgerout <MaskTestName>,<TriggerOutPulse>
MTESt:ONViolation:TRIGgerout? <MaskTestName>
Creates a trigger out pulse on mask violation or successful completion of the test cycle.
Parameters:
<TriggerOutPulse> NOACtion | SUCCess | VIOLation
*RST: NOACtion
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Usage: Asynchronous command
MTESt:ONViolation:RUNexec <MaskTestName>,<RunExecutable>
MTESt:ONViolation:RUNexec? <MaskTestName>
Starts an external application if the command is set to VIOLation or SUCCess.
Use the following commands to set up the application:
EXECutable:NAME on page 1106
EXECutable:PARameter on page 1106
EXECutable:WDIRectory on page 1106
Parameters:
<RunExecutable> NOACtion | SUCCess | VIOLation
See Chapter 18.4.5, "Event parameter", on page 1075
*RST: NOACtion
Parameters for setting and query:
<MaskTestName> String with the name of the mask test
Mask testing

Table of Contents

Other manuals for Rohde & Schwarz RTE Series

Related product manuals