Remote control commands
R&S
®
RTE
1363User Manual 1326.1032.02 ─ 20
SEARch:TRIGger:SLEWrate:TIME <SearchName>,<Time>
SEARch:TRIGger:SLEWrate:TIME? <SearchName>
For the ranges "Within" and "Outside", the slew rate defines the center of a range
which is defined by the limits "±Delta".
For the ranges "Shorter" and "Longer", the slew rate defines the maximum and mini-
mum slew rate limits, respectively.
The range is defined using SEARch:TRIGger:SLEWrate:RANGe.
Parameters:
<Time> Range: 100E-12 to 864
Increment: 100E-9
*RST: 100E-12
Default unit: s
Parameters for setting and query:
<SearchName> String with the name of the search
Usage: Asynchronous command
SEARch:TRIGger:LEVel:TRANsition:LOWer <SearchName>, <SignalSource>,
<Value>
SEARch:TRIGger:LEVel:TRANsition:LOWer? <Key>, <SignalSource>
SEARch:TRIGger:LEVel:TRANsition:UPPer <SearchName>, <SignalSource>,
<Value>
SEARch:TRIGger:LEVel:TRANsition:UPPer? <Key>, <SignalSource>
Set the lower and upper voltage thresholds, respectively. When the signal crosses this
level, the slew rate measurement starts or stops depending on the selected slope.
Parameters:
<Value> Voltage value
Parameters for setting and query:
<SearchName> String with the name of the search
<SignalSource> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | TRK1 | TRK2 |
TRK3 | TRK4 | TRK5 | TRK6 | TRK7 | TRK8 | Z1V1 | Z1V2 |
Z1V3 | Z1V4 | Z1I1 | Z1I2 | Z1I3 | Z1I4 | Z2V1 | Z2V2 | Z2V3 |
Z2V4 | Z2I1 | Z2I2 | Z2I3 | Z2I4
Source of the search, see Chapter 18.4.2, "Waveform parame-
ter", on page 1074
Usage: Asynchronous command