Remote control commands
R&S
®
RTE
1459User Manual 1326.1032.02 ─ 20
BUS<m>:SPI:WSIZe <WordLength>
Sets the number of bits in a message.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<WordLength> Number of bits
Range: 4 to 32
Increment: 1
*RST: 8
Usage: Asynchronous command
BUS<m>:SPI:SCLK:SOURce <SCLKSource>
Sets the input channel of the clock line.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<SCLKSource> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
data, clock and slave select lines.
See Chapter 18.4.2, "Waveform parameter", on page 1074
*RST: C2W1
Usage: Asynchronous command
BUS<m>:SPI:SSELect:SOURce <SlaveSelectSource>
Sets the input channel of the slave select line.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<SlaveSelectSource> NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 |
D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
data, clock and slave select lines.
See Chapter 18.4.2, "Waveform parameter", on page 1074
*RST: None
Protocols