EasyManua.ls Logo

Rohde & Schwarz RTE Series - Page 1461

Rohde & Schwarz RTE Series
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1461User Manual 1326.1032.02 ─ 20
BUS<m>:SPI:MOSI:SOURce <MOSISource>
Sets the input channel of the MOSI line.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<MOSISource> NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 |
D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
data, clock and slave select lines.
See Chapter 18.4.2, "Waveform parameter", on page 1074
*RST: C1W1
Usage: Asynchronous command
BUS<m>:SPI:MOSI:POLarity <MOSIPolarity>
Selects whether transmitted data is high active (high = 1) or low active (low = 1).
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<MOSIPolarity> ACTLow | ACTHigh
*RST: ACTHigh
Usage: Asynchronous command
BUS<m>:SPI:TECHnology <Technology>
Sets the threshold voltage clock, slave select and data lines as defined for various sig-
nal technologies.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Technology> V15 | V25 | V165 | V125 | V09 | VM13 | V38 | V20 | V0 | MANual
V15 | V25 | V165 | V125 | V09 | V38 | V20 | V0
1.5 V, 2.5 V, 1.65 V ... respectively
VM13
-1.3 V (negative value)
MANual
Manual setting of user-defined values with
BUS<m>:SPI:SCLK|SSEL|MISO|MOSI:THReshold.