Remote control commands
R&S
®
RTE
1461User Manual 1326.1032.02 ─ 20
BUS<m>:SPI:MOSI:SOURce <MOSISource>
Sets the input channel of the MOSI line.
Suffix: 
<m>
.
1..4
Selects the serial bus.
Parameters:
<MOSISource> NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 |
D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Digital and analog channels cannot be used at the same time for
data, clock and slave select lines.
See Chapter 18.4.2, "Waveform parameter", on page 1074
*RST:  C1W1
Usage:  Asynchronous command
BUS<m>:SPI:MOSI:POLarity <MOSIPolarity>
Selects whether transmitted data is high active (high = 1) or low active (low = 1).
Suffix: 
<m>
.
1..4
Selects the serial bus.
Parameters:
<MOSIPolarity> ACTLow | ACTHigh
*RST:  ACTHigh
Usage:  Asynchronous command
BUS<m>:SPI:TECHnology <Technology>
Sets the threshold voltage clock, slave select and data lines as defined for various sig-
nal technologies.
Suffix: 
<m>
.
1..4
Selects the serial bus.
Parameters:
<Technology> V15 | V25 | V165 | V125 | V09 | VM13 | V38 | V20 | V0 | MANual
V15 | V25 | V165 | V125 | V09 | V38 | V20 | V0
1.5 V, 2.5 V, 1.65 V ... respectively
VM13
-1.3 V (negative value)
MANual
Manual setting of user-defined values with
BUS<m>:SPI:SCLK|SSEL|MISO|MOSI:THReshold.
Protocols