Remote control commands
R&S
®
RTE
1658User Manual 1326.1032.02 ─ 20
<n> *
Return values:
<FrameType> MAC | IDLE | SLEep | EOS | UNKNown
*RST: MAC
Usage: Query only
BUS<m>:ETHernet:WORD<n>:STATe?
Returns the frame state of the selected word in the current acquisition.
Suffix:
<m>
.
1..4
<n> *
Return values:
<State> OK | ERR_PREAMBLE | ERR_LENGTH | UNCorrelated |
INComplete
OK
No error detected
ERR_PREAMBLE
Error in the preamble of the selected word
ERR_SFD
Error in the start frame delimiter (SFD). The value of a correct
SFD byte is 171. The SFD is transmitted LSB first.
ERR_LENGTH
Error in the number of bits in the selected word
*RST:
OK
Usage: Query only
BUS<m>:ETHernet:WORD<n>:STARt?
Returns the frame start time of the selected word in the current acquisition.
Suffix:
<m>
.
1..4
<n> *
Return values:
<Start> Range: -100E+24 to 100E+24
Increment: 100E-12
Default unit: s
Usage: Query only
BUS<m>:ETHernet:WORD<n>:STOP?
Returns the frame stop time of the selected word in the current acquisition.