Remote control commands
R&S
®
RTE
1798User Manual 1326.1032.02 ─ 20
MD
Sets the trigger to the MDATA field (high-speed data packet PID
for split and high bandwidth isochronous transactions).
*RST: D0
TRIGger<m>:USB:DCONdition <OperatorData>
Sets the operator (equal or unequal) to set a specific payload data pattern.
Parameters:
<OperatorData> OFF | ANY | EQUal | NEQual
*RST: EQUal
TRIGger<m>:USB:DPOPerator
<DataPosOperator>
Sets the operator (any or equal) for the payload data index position.
Parameters:
<DataPosOperator> ANY | OFF | EQUal
ANY = OFF
The position of the bit pattern within the payload data is not rele-
vant.
EQUal
Sets the operator for specifying a special start position for the bit
pattern within the payload data.
*RST: ANY
TRIGger<m>:USB:DPOSition <DataPosition>
Specifies the position within a payload data packet, in which a special data pattern is to
be searched.
Parameters:
<DataPosition> Range: 0 to 1024
Increment: 1
*RST: 0
TRIGger<m>:USB:ECONdition <EndpointOptor>
Sets the operator to set a specific endpoint or an endpoint range. The endpoint values
are set with TRIGger<m>:USB:EMIN and TRIGger<m>:USB:EMAX.
Parameters:
<EndpointOptor> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange