Remote control commands
R&S
®
RTE
1961User Manual 1326.1032.02 ─ 20
POWer:SLEWrate:RESult:PPEak?.................................................................................1962
POWer:SLEWrate:RESult:RMS?................................................................................... 1962
POWer:SLEWrate:RESult:STDDev?.............................................................................. 1963
POWer:SLEWrate:RESult:WFMCount?.......................................................................... 1963
POWer:SLEWrate:SOURce <Source>
Selects the source for the slew rate measurement.
Parameters:
<Source> CURRENT | VOLTAGE
*RST: VOLTAGE
POWer:SLEWrate:AUTO
<Autoscale>
Selects the scaling for the display of the results.
Parameters:
<Autoscale> AUTO | MANUAL
*RST: MANUAL
POWer:SLEWrate:AVGDeriv <AvgDerivative>
Activates or deactivates average.
Parameters:
<AvgDerivative> ON | OFF
*RST: ON
POWer:SLEWrate:EXECute
Starts the slew rate measurement.
Usage:
Event
ACQuire:ARESet:MODE <ArtmRst>
Defines when the envelope and average evaluation restarts.
Parameters:
<ArtmRst> NONE | TIME | WFMS
TIME
Restarts the envelope and average calculation after the time
defined with ACQuire:ARESet:TIME.
WFMS
Restarts the envelope and average calculation after a number of
acquired waveforms defined with ACQuire:ARESet:COUNt
on page 1962.
*RST: NONE
Power analysis (option R&S RTE-K31)