Triggers
R&S
®
RTE
260User Manual 1326.1032.02 ─ 20
Most video signals have an output impedance of 75. The channel inputs of the
R&S RTE have an input impedance of 50 or 1 M. Make sure to provide the ade-
quate matching to ensure amplitude fidelity. A simple 75 feed-through termination
combined with 1 M oscilloscope inputs is suitable for most applications.
Once the trigger is set correctly, you can use cursor and automatic measurements to
perform amplitude and timing measurements.
Figure 6-1: Trigger on all lines of a PAL signal with positive signal polarity, trigger level = 300 mV
Standard
Selects the TV standard or "Custom" for user-defined signals.
Trigger types

Table of Contents