Waveform analysis
R&S
®
RTE
298User Manual 1326.1032.02 ─ 20
Mode ← Arithmetic
Waveform arithmetic builds the resulting waveform from several consecutive acquisi-
tions and subsequent math calculations of the signal. For details, see "Arithmetic"
on page 182.
"Original"
The original results are displayed.
"Envelope"
The envelope curve of all acquired and calculated results is dis-
played.
"Average"
The average of all acquired and calculated results is displayed.
"RMS"
The root mean square of the math data is displayed. The result is the
average power spectrum. If you measure the channel power on this
RMS spectrum, you get the same result as for the average channel
power measurement on waveforms.
"MinHold"
Determines the minimum result for each input value from the data of
the current acquisition and some acquisitions before.
"MaxHold"
Determines the maximum result for each input value from the data of
the current acquisition and some acquisitions before.
Remote command:
CALCulate:MATH<m>:ARIThmetics on page 1228
Reset ← Arithmetic
Forces the immediate restart of the envelope and average calculation for all wave-
forms.
Remote command:
ACQuire:ARESet:IMMediate on page 1115
Average count (N-single count) ← Arithmetic
Access:
TRIGGER > "Control" tab > "Average count (N-single count)"
[ACQUISITION] > "Average count (N-single count)"
[HORIZONTAL] > "Fast Segmentation" tab > disable "Acquire maximum" >
"Required"
[MATH] > "Setup" tab > "Mode" is not "Off" > "Average count"
The acquisition and average count has several effects:
It sets the number of waveforms acquired with [RUN N× SINGLE]
It defines the number of waveforms used to calculate the average waveform.
Thus, the instrument acquires sufficient waveforms to calculate the correct average
if "Average" is enabled for waveform arithmetic. The higher the value is, the better
the noise is reduced.
It sets the number of acquisitions to be acquired in a fast segmentation acquisition
series. Thus, you can acquire exactly one fast segmentation acquisition series with
[RUN N× SINGLE].
If fast segmentation is enabled and configured to acquire the maximum number of
acquisitions, the acquisition count is set to that maximum number and cannot be
changed.
See also: "Number of acquisitions" on page 184.
It is the "Finished" criteria for the state of a mask test.