Waveform analysis
R&S
®
RTE
314User Manual 1326.1032.02 ─ 20
If a history replay is running, the field shows the number of the currently shown acquisi-
tion.
Remote command: 
CHANnel<m>[:WAVeform<n>]:HISTory:CURRent on page 1232
Start acq
Sets the index of the first (oldest) acquisition to be displayed or exported. The index is
always negative.
The number of stored history acquisitions is shown in Available acquisitions on the
"Information" tab.
Remote command: 
CHANnel<m>[:WAVeform<n>]:HISTory:STARt on page 1232
Stop acq
Sets the index of the last (newest) acquisition to be displayed or exported. The newest
acquisition of the complete acquisition series always has the index "0".
Remote command: 
CHANnel<m>[:WAVeform<n>]:HISTory:STOP on page 1233
Select all
All acquisitions that are saved in the memory are in the viewer.
Current
Sets the newest acquisition in the sample memory as "Stop acq" and "Current acq".
This acquisition always has the index "0".
Oldest
Sets the oldest acquisition in the sample memory as "Start acq" and "Current acq".
Auto repeat
If selected, the replay of the history waveform sequence repeats automatically. Other-
wise, the replay stops at the "Stop acq".
Remote command: 
CHANnel<m>[:WAVeform<n>]:HISTory:REPLay on page 1234
Play
Starts and stops the replay of the history waveforms from "Start acq" to "Stop acq".
Remote command: 
CHANnel<m>[:WAVeform<n>]:HISTory:PLAY on page 1233
Replay time per acq.
Sets the display time for one acquisition. The shorter the time, the faster the replay is.
The setting takes effect for history replay and the display of a Fast Segmentation ser-
ies, see Chapter 5.2.3, "Fast segmentation", on page 183.
Remote command: 
CHANnel<m>[:WAVeform<n>]:HISTory:TPACq on page 1233