Measurements
R&S
®
RTE
384User Manual 1326.1032.02 ─ 20
Figure 8-11: Long-term measurement with statistics
1 = +Peak, maximum
2 = Average + standard deviation
3 = Average
2 = Average - standard deviation
1 = -Peak, minimum
Stopping and starting the acquisition does not reset statistics but only stops and contin-
ues them.
The instrument only resets statistical evaluation if you change measurement setup:
●
Select measurement type
●
Create or modify gate
●
Enable/disable long-term measurement and histogram
●
Enable continuous autoscale with enabled histogram
●
Switch on/off channels
●
Enable/disable cursors
●
Tap "Reset" or "Clear sreen results"
After a reset, new statistics are compiled beginning with the next acquired waveform.
If limit and margin check is enabled, the icons in the result table indicate if statistical
results exceed a limit or margin. These violations do not initiate an action.