Protocol analysis
R&S
®
RTE
829User Manual 1326.1032.02 ─ 20
See "End Point" on page 832
See "Address" on page 832
SOF ← Token
For the trigger token type "SOF", the following conditions can be set:
You can refine the trigger condition:
See "PID error " on page 832
See "Frame number" on page 833
Data
Sets the trigger to one out of four different data types:
"DATA0"
"DATA1"
"DATA2"
"MDATA"
If no additional conditions are set, the trigger instant is after the PID.
If the PID error check is selected:
If an error is found in the PID's complementary form, the trigger is set immediately
after the 8
th
bit of the PID
If no PID error is found, no trigger is set, even if other selected conditions are met.
Otherwise, the trigger instant is defined by the first occurrence of any of the specified
additional conditions.
You can refine the trigger condition:
See "PID error " on page 832
USB (option R&S
RTE-K60)