Instrument Function
R&S
®
SMU200A
249Operating Manual 1007.9845.32 ─ 15
Note: Activating the RF sweep automatically deactivates the list mode, level sweep and
LF sweep.
In case of two-path instrument, this affects the respective path.
SCPI command:
[:SOURce<hw>]:FREQuency:MODE on page 714
Mode - Frequency Sweep
Selects the Sweep instrument operating mode and the Sweep mode.
"Auto"
Sets an automatic repeated sweep cycle. If a different sweep mode was
activated prior to the "Auto" mode, the cycle continues from the current
sweep setting.
The "Reset Sweep" button resets the sweep to the start frequency.
Example:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR AUTO
SOUR:FREQ:MODE SWE
"Single"
Sets a single sweep cycle. The sweep is triggered by the "Execute Single
Sweep" button.
If a different sweep mode was activated prior to the "Single" mode, the
current sweep is stopped. The "Single" sweep always starts at the start
frequency when triggered.
The "Reset Sweep" button resets the sweep to the start frequency.
Example:
SOUR:SWE:FREQ:MODE AUTO
TRIG:FSW:SOUR SING
SOUR:FREQ:MODE SWE
SOUR:SWE:FREQ:EXEC
RF Signal and Analog Modulations - RF Block