Instrument Function
R&S
®
SMU200A
256Operating Manual 1007.9845.32 ─ 15
"Single"
Sets a single sweep cycle. The sweep is triggered by the "Execute Single
Sweep" button.
If a different sweep mode was activated prior to the "Single" mode, the
current sweep is stopped. The "Single" sweep always starts at the start
level.
The "Reset Sweep" button resets the sweep to the start level.
Example:
SOUR:SWE:POW:MODE AUTO
TRIG:PSW:SOUR SING
SOUR:POW:MODE SWE
SOUR:SWE:POW:EXEC
"Step"
Sets a step-by-step sweep cycle.
If this mode is activated, the cursor moves to the value displayed for
"Current Level". Each sweep step is triggered by a variation of the value
in the "Current Level" entry window. The step width is set below at entry
field "Step".
If this mode is activated, the cursor moves to the value displayed for
"Current Level". If a different sweep mode was activated prior to the
"Step" mode, the current sweep is stopped. The "Step" sweep starts at
the current level value.
The "Reset Sweep" button resets the sweep to the start level.
Example:
SOUR:SWE:POW:MODE MAN
SOUR:SWE:POW:STEP 0.5
SOUR:POW:MODE SWE
SOUR:POW:MAN -16
The value entered with command SOUR:SWE:POW:STEP sets the step
width.
The value entered with command SOUR:POW:MAN has no effect, the
command only triggers the next sweep step. However, the value has to
be in the currently set sweep range (start to stop). In remote control only
a step-by-step sweep from start to stop frequency is possible.
RF Signal and Analog Modulations - RF Block