Instrument Function
R&S
®
SMU200A
307Operating Manual 1007.9845.32 ─ 15
SCPI command:
[:SOURce<hw>]:AWGN:POWer:NOISe on page 576
Noise Level (Total Bandwidth) - AWGN
Displays the level of the noise signal in the total bandwidth in "Noise Only" and "Additive
Noise" mode.
SCPI command:
[:SOURce<hw>]:AWGN:POWer:NOISe:TOTal on page 576
Carrier + Noise Level / Signal + Noise Level (System Bandwidth) / Carrier + Inter-
ferer Level / Signal + Interferer Level
(Additive Noise and CW Interferer)
Displays the overall level of the noise/interferer signal plus useful signal.
Tip: In the mode "Additive Noise" the "Display Output Results For" (analog or digital) has
to be set before.
The name of this parameter depends on the selected "Display Mode".
SCPI command:
[:SOURce<hw>]:AWGN:POWer:SUM on page 577
Carrier + Noise PEP / Signal + Noise PEP (Total Bandwidth) / Carrier + Interferer
PEP / Signal + Interferer PEP
(Additive Noise and CW Interferer)
Displays the peak envelope power of the overall signal comprised of noise signal plus
useful signal.
The name of this parameter depends on the selected "Display Mode".
Note: The peak envelope power (PEP) specified in the header corresponds to the PEP
value of the carrier.
In the mode "Additive Noise" the "Display Output Results For" (analog or digital) has to
be set before.
SCPI command:
[:SOURce<hw>]:AWGN:POWer:SUM:PEP on page 577
Impairment of Digital I/Q Signal and Noise Generator - AWGN/IMP Block