Remote Control Commands
R&S
®
SMU200A
576Operating Manual 1007.9845.32 ─ 15
[:SOURce<hw>]:AWGN:POWer:NOISe <Noise>
Available for modes Additive Noise, Noise Only and CW Interferer (SOUR:AWGN:MODE
ADD|ONLY|CW).
This command sets or queries the noise level in the system bandwidth depending on the
selected modes:
●
"Additive Noise" and "CW Interferer" mode (SOUR:AWGN:MODE ADD|CW)
- "Reference Mode Carrier" (SOUR:AWGN:POW:RMOD CARR)
The command queries the noise/interferer level which is derived from the entered
C/N value. The carrier level is set with command SOUR:AWGN:POW:CARRier.
- "Reference Mode Noise" (SOUR:AWGN:POW:RMOD NOIS
The command sets the noise/interferer level. The level of the carrier signal is
derived from the entered C/N value.
●
"Noise Only" mode
The command sets the noise level.
Parameters:
float
dBm
<Noise>
Default unit:
SOUR:AWGN:POW:NOIS?
queries the noise level in the system bandwidth.
Response: 10
the noise level in the system bandwidth is 10 dBm.
Example:
[:SOURce<hw>]:AWGN:POWer:NOISe:TOTal?
Available for modes Additive Noise and Noise Only (SOUR:AWGN:MODE ADD|ONLY)
This command queries the noise level in the total bandwidth.
Tip: "Display Output Results For" ([:SOURce<hw>]:AWGN:DISP:ORESults) the ana-
log or digital signal path has to be set before.
Return values:
float<Total>
SOUR:AWGN:POW:NOIS:TOT?
queries the noise level in the total bandwidth.
Response: 15
the noise level in the total bandwidth is 15 dBm.
Example:
Query onlyUsage:
[:SOURce<hw>]:AWGN:POWer:RMODe <Rmode>
This command selects the mode for setting the interfering signal.
This command is available for modes "Additive Noise" and "CW Interferer"
(SOUR:AWGN:MODE ADD|CW) and Display Mode set to RF (AWGN:DISP:MODE RF).
SOURce Subsystem