Remote Control Commands
R&S
®
SMU200A
575Operating Manual 1007.9845.32 ─ 15
Parameters:
float
min to 20
0.01
-30
dBm
<Carrier>
Range:
Increment:
*RST:
Default unit:
AWGN:DISP:MODE RF
sets the display mode to RF
AWGN:POW:CARR?
queries the carrier level
Example:
[:SOURce<hw>]:AWGN:POWer:MODE <Mode>
This command selects the mode for setting the noise level.
This command is available for mode "Additive Noise" (SOUR:AWGN:MODE ADD).
Parameters:
CN|SN | EN
CN|SN
available for mode Additive Noise (SOUR:AWGN:MODE ADD).
The noise level is set on the basis of the value entered for the
carrier/noise or signal/noise ratio (AWGN:CNR|SNR).
Whether a selection of C/N or S/N is enabled, depends on the
selected "Display Mode" (AWGN:DISP:MODE BB|RF)
EN
The noise level is set on the basis of the value entered for the ratio
of bit energy to noise power density (AWGN:ENR).
CN|SN
<Mode>
*RST:
AWGN:DISP:MODE RF
sets the display mode to RF
SOUR:AWGN:POW:MODE CN
the noise level is set on the basis of the value entered for the car-
rier/noise ratio (AWGN:CNR).
Example:
SOURce Subsystem