EasyManua.ls Logo

Rohde & Schwarz SMU200A - Page 462

Rohde & Schwarz SMU200A
888 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Function
R&S
®
SMU200A
442Operating Manual 1007.9845.32 ─ 15
Indicates the peak level acquired with "Auto Level Set".
SCPI command:
[:SOURce]:BBIN:POWer:PEAK on page 695
Level
Indicates the estimated rms level acquired with "Auto Level Set".
SCPI command:
[:SOURce]:BBIN:POWer:RMS on page 695
Impairements
For analog input signals the section "Impairments" is available in the "Baseband Input
Settings" dialog. Additionally to the internal calibration of the instrument a DC offset, gain
imbalance and time delay can be set to I and Q signal components. The section Signal
Monitoring indicates the overload of the A/D - Converter.
Gain Imbalance (Q versus I)
Analog Input only
Enables to amplify the Q component of the input signal additionally to the internal cali-
bration. This setting is needed e.g. to balance the I and Q signal channel by exactly the
same degree.
SCPI command:
[:SOURce<hw>]:BBIN:GIMBalance on page 692
I Offset
Analog Input only
Sets a DC offset to the I component of the input signal additionally to the internal cali-
bration. This setting is needed to compensate an offset that has been fed in from the
external signal.
SCPI command:
[:SOURce<hw>]:BBIN:OFFSet:I on page 693
Q Offset
Analog Input only
Sets a DC offset to the Q component of the input signal additionally to the internal cali-
bration. This setting is needed to compensate an offset that has been fed in from the
external signal.
SCPI command:
[:SOURce<hw>]:BBIN:OFFSet:Q on page 693
External Baseband Signal - Baseband Input

Table of Contents

Other manuals for Rohde & Schwarz SMU200A