Remote Control Commands
R&S
®
SMU200A
590Operating Manual 1007.9845.32 ─ 15
Parameters:
INTernal|EXTernal | BEXTernal|OBASeband
INTernal
Triggering is executed by means of the Trigger command in the
case of remote control, and by means of "Execute Trigger" in the
case of manual operation.
EXTernal
Triggering is executed by means of the signal on the TRIGGER 1
connector.
BEXTernal
Triggering is executed by means of the signal on the TRIGGER 2
connector.
OBASeband
Triggering is executed by means of the signal from the second
path.
INTernal
<Source>
*RST:
BB:ARB:TRIG:SOUR INT
sets internal triggering.
Example:
[:SOURce<hw>]:BB:ARB:TRIGger[:EXTernal<1|2>]:DELay <Delay>
The command specifies the trigger delay (expressed as a number of samples) for external
triggering.
The numeric suffix to EXTernal distinguishes between the external trigger via the TRIG-
GER 1 (suffix 1) and TRIGGER 2 (suffix 2) connector.
Parameters:
float
0 to 2^32-1
1
0
sample
<Delay>
Range:
Increment:
*RST:
Default unit:
BB:ARB:TRIG:SOUR EXT
selects an external trigger via the TRIGGER 1 connector
BB:ARB:TRIG:DEL 200
sets a delay of 200 samples for the trigger.
Example:
[:SOURce<hw>]:BB:ARB:TRIGger[:EXTernal<1|2>]:INHibit <Inhibit>
The command specifies the number of samples by which a restart is to be inhibited fol-
lowing a trigger event. This command applies only in the case of external triggering.
The numeric suffix to EXTernal distinguishes between the external trigger via the TRIG-
GER 1 (suffix 1) and TRIGGER 2 (suffix 2) connector.
SOURce Subsystem