Remote Control Commands
R&S
®
SMU200A
596Operating Manual 1007.9845.32 ─ 15
Parameters:
float
0.01 Hz
1 kHz
<Frequency>
Increment:
*RST:
BB:ARB:TSIG:RECT:FREQ 100 kHz
sets the frequency of the test signal of 100 kHz.
Example:
[:SOURce<hw>]:BB:ARB:TSIGnal:RECTangle:OFFSet <Offset>
Sets the DC component.
Parameters:
float
-1.0 to 1.0
0.1
1 FS
FS
<Offset>
Range:
Increment:
*RST:
Default unit:
BB:ARB:TSIG:RECT:OFFS 0.5
sets the DC component
Example:
[:SOURce<hw>]:BB:ARB:TSIGnal:RECTangle:SAMPles <Samples>
Sets the number of sample values required for the rectangular signal per period.
Parameters:
float
4 to 1000
1 sample
<Samples>
Range:
Increment:
BB:ARB:TSIG:RECT:SAMP 400
sets 400 samples per period
Example:
[:SOURce<hw>]:BB:ARB:TSIGnal:SINE:FREQuency <Frequency>
The command sets the frequency of the simple sinusoidal test signal. This signal is used
as output via the I channel. A sine wave of the same frequency but optionally phase-
shifted is generated on the Q path (ARB:TSIGnal:SINE:POFFset).
Parameters:
float
0.01 Hz
1 kHz
Hz
<Frequency>
Increment:
*RST:
Default unit:
BB:ARB:TSIG:SINE:FREQ 100 kHz
sets a sine signal of 100 kHz.
Example:
SOURce Subsystem