Remote Control Commands
R&S
®
SMU200A
605Operating Manual 1007.9845.32 ─ 15
BB:ARB:MCAR:EDIT:CARR:STAR 4
the carrier range starts at carrier 2.
BB:ARB:MCAR:EDIT:CARR:STOP 20
the carrier range stops at carrier 20.
BB:ARB:MCAR:EDIT:CARR:STAT ON
sets all the carriers in the carrier range (2 to 20) to ON.
BB:ARB:MCAR:EDIT:CARR:EXEC
transfers the assistant settings for carrier 2 to 20 into the carrier
table.
Example:
EventUsage:
[:SOURce<hw>]:BB:ARB:MCARrier:EDIT:CARRier:FILE <File>
Parameters:
string<File>
BB:ARB:MCAR:EDIT:CARR:FILE 'D:/IQ_wcdma'
selects input file IQ_wcdma. The data of the file are modulated
onto the carriers in the defined carrier range.
Example:
[:SOURce<hw>]:BB:ARB:MCARrier:EDIT:CARRier:PHASe:STEP <Step>
The command sets the step width by which the start phases of the carriers in the defined
carrier range will be incremented.
The phase settings are only valid if optimization of the crest factor is disabled
(:SOURce:BB:ARB:MCARrier:CFACtor:MODE OFF).
Parameters:
float
0.0 to 360.0
0.01
0
DEG
<Step>
Range:
Increment:
*RST:
Default unit:
BB:ARB:MCAR:EDIT:CARR:PHAS 90 DEG
sets a start phase of 90° for the carriers in the carrier range.
BB:ARB:MCAR:EDIT:CARR:PHAS:STEP 1 DEG
the start phase is incremented by 1° for each carrier, i.e. the first
carrier has a start phase of 90°, the second a start phase of 91°,
etc.
Example:
SOURce Subsystem