Remote Control Commands
R&S
®
SMU200A
604Operating Manual 1007.9845.32 ─ 15
EventUsage:
[:SOURce<hw>]:BB:ARB:MCARrier:EDIT:CARRier:DELay:STEP <Step>
The command sets the step width by which the start delays of the carriers in the defined
carrier range will be incremented.
Parameters:
float
-1s to +1s
1ns
0s
s
<Step>
Range:
Increment:
*RST:
Default unit:
BB:ARB:MCAR:EDIT:CARR:DEL 5 us
sets a start delay of 5 us for the carriers in the carrier range.
BB:ARB:MCAR:EDIT:CARR:DEL:STEP 1 us
the start delay is incremented by 1us for each carrier, i.e. the first
carrier has a start delay of 5us, the second a start delay of 6 us,
etc.
Example:
[:SOURce<hw>]:BB:ARB:MCARrier:EDIT:CARRier:DELay[:STARt] <Start>
The command sets the start delay for the individual carriers in the defined carrier range.
If the command :BB:ARB:MCAR:EDIT:CARR:DEL:STEP is used to define a step width,
the delay entered here applies only to the starting carrier. The delays of the remaining
carriers are stepped up or down by the delay value specified in
the :BB:ARB:MCAR:EDIT:CARR:DEL:STEP command.
Parameters:
float
-1s to +1s
1 ns
0 s
s
<Start>
Range:
Increment:
*RST:
Default unit:
BB:ARB:MCAR:EDIT:CARR:DEL 5us
sets a start delay of 5 us for the carriers in the carrier range.
Example:
[:SOURce<hw>]:BB:ARB:MCARrier:EDIT:CARRier:EXECute
The command adopts the settings for the carrier range which has been defined using
the :BB:ARB:MCAR:EDIT:CARR:... commands.
SOURce Subsystem