Remote Control Commands
R&S
®
SMU200A
658Operating Manual 1007.9845.32 ─ 15
[:SOURce<hw>]:BB:GRAPhics:SRATe:USER <User>
(Enabled for BB:GRAPH:SRAT:MODE USER)
Selects the signal bandwidth for which the display is to be generated. The setting range
moves between the minimum and maximum bandwidth which is possible for the selected
display. The selection is made graphically by moving the pointer.
Parameters:
float
0.1 to 100.0
0.1
10
PCT
<User>
Range:
Increment:
*RST:
Default unit:
BB:GRAP:SRAT:MODE USER
sets the sample rate mode.
BB:GRAP:SRAT:USER 20
sets the sample rate factor.
Example:
Example:
[:SOURce<hw>]:BB:GRAPhics:STATe <State>
Activates/deactivates the selected graphical display.
Parameters:
0|1|OFF|ON
OFF
<State>
*RST:
BB:GRAP:STAT ON
activates the graphical graphics.
Example:
[:SOURce<hw>]:BB:GRAPhics:TRIGger:SOURce <Source>
Defines the trigger for the starting time of the graphic recording.
SOURce Subsystem