Remote Control Commands
R&S
®
SMU200A
675Operating Manual 1007.9845.32 ─ 15
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:PHASe[:STARt] <Start>
The command sets the start phase for the individual carriers in the defined carrier range.
If the command :BB:MCCW:EDIT:CARR:PHAS:STEP is used to define a step width, the
phase entered here applies only to the starting carrier. The phases of the remaining car-
riers are stepped up or down by the phase value specified in
the :BB:MCCW:EDIT:CARR:PHAS:STEP command.
The phase settings are only valid if optimization of the crest factor is disabled
(SOURce:BB:MCCW:CFACtor:MODE OFF).
Parameters:
float
0 DEG to 360 DEG
0.01 DEG
0 DEG
<Start>
Range:
Increment:
*RST:
BB:MCCW:EDIT:CARR:PHAS 90 DEG
sets a start phase of 90° for the carriers in the carrier range.
Example:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:POWer:STEP <Step>
The command sets the step width by which the starting power of the carriers in the defined
carrier range will be incremented.
Parameters:
float
-80 dB to 80 dB
0.01 dB
0 dB
<Step>
Range:
Increment:
*RST:
BB:MCCW:EDIT:CARR:POW -80dB
sets a power of -80 dB for the carriers in the carrier range.
BB:MCCW:EDIT:CARR:POW:STEP 1 dB
the power phase is incremented by dB for each carrier, i.e. the
first carrier has -80 dB, the second -79 dB, etc.
Example:
[:SOURce<hw>]:BB:MCCW:EDIT:CARRier:POWer[:STARt] <Start>
The command sets the power for the individual carriers in the defined carrier range. If the
command :BB:MCCW:EDIT:CARR:POW:STEP is used to define a step width, the power
entered here applies only to the starting carrier. The power of the remaining carriers is
stepped up or down by the power specified in the :BB:MCCW:EDIT:CARR:POW:STEP
command.
SOURce Subsystem