Remote Control Commands
R&S
®
SMU200A
737Operating Manual 1007.9845.32 ─ 15
Parameters:
PEP|LEVel
PEP
The signal level is expressed in terms of a peak envelope power
value (PEP). The level entry fields and the corresponding units are
displayed in the status bar.
LEVel
The signal level is expressed in terms of a rms value. The level
entry field and the corresponding units are displayed in the status
bar.
LEVel
<Via>
*RST:
IQ:OUTP:POW:VIA PEP
the I/Q-signal level is set via PEP for the signal of path A.
Example:
[:SOURce<hw>]:IQ:OUTPut:DIGital:SRATe <Srate>
Using the sample rate source User the command sets the sample rate value of the output
signal. This command is not active for external clock sources
Parameters:
float
400 Hz to 100 MHz
0.01 Hz
100 MHz
<Srate>
Range:
Increment:
*RST:
IQ:OUTP:DIG:SRAT:MODE USER
with sample rate source User the internally generated clock ref-
erence signal is selected.
IQ:OUTP:DIG:SRAT 100 MHz
the sample rate for the output signal is set to 100 MHz.
Example:
B13 and B18Options:
[:SOURce<hw>]:IQ:OUTPut:DIGital:SRATe:FIFO[:STATus]?
This command queries the state of the signal transmission.
Return values:
OK|OFLow|URUN
OK
Resampling is working properly.
OFLow (Overflow)
The external sample rate frequency is too high which causes an
overflow in signal transmission.
URUN (Underrun)
The external sample rate is too low which causes an underrun in
signal transmission.
<Status>
SOURce Subsystem