Remote Control Commands
R&S
®
SMU200A
769Operating Manual 1007.9845.32 ─ 15
Parameters:
CW|FIXed|SWEep
CW|FIXed
The instrument operates at a constant level. CW and FIXed are
synonyms. The output level is set with :SOURce:POWer.
SWEep
The instrument operates in SWEep mode.
The level is set using the commands SOURce:POWer:STARt;
STOP; MANual.
LIST
The instrument processes a list of frequency and level settings.
The List mode settings are made in the SOURce:LIST subsystem.
The setting SOURce:POWer:MODE LIST also sets the command
SOURce:FREQuency:MODE automatically to LIST.
CW
<Mode>
*RST:
POW:MODE SWEep
sets the SWEep mode. The settings under SOURce:POW:STARt;
STOP; MANual become effective.
Example:
SCPI conformUsage:
[:SOURce<hw>]:POWer:PEP?
With digital modulation and all digital standards, the command returns the peak envelope
power (PEP) in dBm (the value which is shown under the Level specification in the PEP
lines in the display header).
Return values:
float<Pep>
POW:PEP?
queries the PEP value for RF output A in the case of digital mod-
ulation.
Response: 4
the PEP value for RF output A is 4 dBm.
Example:
Query only
SCPI conform
Usage:
SOURce Subsystem