Remote Control Commands
R&S
®
SMU200A
768Operating Manual 1007.9845.32 ─ 15
[:SOURce<hw>]:POWer:MANual <Manual>
In Sweep mode (:SOUR:POW:MODE SWE) the command sets the level for the next
sweep step in the Step sweep mode (:SOUR:SWE:POW:MODE MAN). Here only level
values between the settings [:SOUR]:POW:STAR and ...:STOP are permitted. Each
sweep step is triggered by a separate :SOUR:POW:MAN command.
As with the "Level" value entered in the "RF Level" menu, the OFFSet value is also taken
into consideration with this command.
The specified value range is therefore only effective if :SOURce:POWer:OFFSet is set
to 0. The value range for other OFFset values can be calculated using the following
formula:
Minimum level + OFFSet ... Maximum level + OFFSet
Parameters:
float
Minimum level ... Maximum level
The value range for the level setting varies according to the
instrument model
The value range depends on whether or not the instrument is
equipped with the option High Power.
The values are given in the data sheet.
0.01 dBm
-30 dBm
dBm
<Manual>
Increment:
*RST:
Default unit:
POW:SWE:MODE MAN
sets the Step sweep mode for RF output A.
POW:MAN -5 dBm
sets an RF level of -5 dBm for the next setting in the Step sweep
mode for RF output A.
POW:MODE SWE
sets the Level Sweep mode for RF output A.
POW:MAN -5.5 dBm
triggers the next sweep step with a level of -5.5 dBm.
Example:
SCPI conformUsage:
[:SOURce<hw>]:POWer:MODE <Mode>
The command sets the instrument operating mode and therefore also the commands
used to set the output level.
POW:MODE LIST sets FREQ:MODE LIST
In case of two-path instrument, this affects the selected path.
SOURce Subsystem