Remote Control Commands
R&S
®
SMU200A
785Operating Manual 1007.9845.32 ─ 15
STATus:OPERation:CONDition? <Condition >
The command queries the content of the CONDition part of the STATus:OPERation reg-
ister. This part contains information on the action currently being performed in the instru-
ment. The content is not deleted after being read out since it indicates the current hard-
ware status.
Parameters:
string<Condition >
STAT:OPER:COND
queries the Status:Operation:Condition register.
Example:
Query only
SCPI conform
Usage:
STATus:OPERation:ENABle <Enable>
The command sets the bits of the ENABle part of the STATus:OPERation register. This
setting determines which events of the Status-Event part are forwarded to the sum bit in
the status byte. These events can be used for a service request.
Parameters:
string<Enable>
STAT:OPER:ENAB 32767
all events are forwarded to the sum bit of the status byte.
Example:
SCPI conformUsage:
STATus:OPERation[:EVENt]? <EVENt>
The command queries the content of the EVENt part of the STATus:OPERation register.
This part contains information on the actions performed in the instrument since the last
readout. The content of the EVENt part is deleted after being read out.
Parameters:
string<EVENt>
STAT:OPER:EVEN
queries the STATus:OPERation:EVENt register.
Example:
Query only
SCPI conform
Usage:
STATus:OPERation:PTRansition <Ptransition>
The command sets the bits of the PTRansition part of the STATus:OPERation register.
If a bit is set, a transition from 0 to 1 in the condition part causes an entry to be made in
the EVENt part of the register. A new event in the hardware is thus registered, e.g. the
start of an adjustment.