Remote Control Commands
R&S
®
SMU200A
787Operating Manual 1007.9845.32 ─ 15
STAT:OPER:COND
queries the Status:Questionable:Condition register.
Example:
Query only
SCPI conform
Usage:
STATus:QUEStionable:ENABle <Enable>
The command sets the bits of the ENABle part of the STATus:QUEStionable register.
This setting determines which events of the Status-Event part are enabled for the sum
bit in the status byte. These events can be used for a service request.
Parameters:
string<Enable>
STAT:OPER:ENAB 1
problems when performing an adjustment cause an entry to be
made in the sum bit.
Example:
SCPI conformUsage:
STATus:QUEStionable:NTRansition <Ntransition>
The command sets the bits of the NTRansition part of the STATus:QUEStionable register.
If a bit is set, a transition from 1 to 0 in the condition part causes an entry to be made in
the EVENt part of the register.
Parameters:
string<Ntransition>
STAT:OPER:NTR 0
A transition from 1 to 0 in the condition part of the Status:Ques-
tionable register does not cause an entry to be made in the EVENt
part.
Example:
SCPI conformUsage:
STATus:QUEStionable:PTRansition <Ptransition>
The command sets the bits of the NTRansition part of the STATus:QUEStionable register.
If a bit is set, a transition from 1 to 0 in the condition part causes an entry to be made in
the EVENt part of the register.
Parameters:
string<Ptransition>
STAT:OPER:PTR 32767
All transitions from 0 to 1 in the condition part of the Status:Ques-
tionable register cause an entry to be made in the EVENt part.
Example:
SCPI conformUsage:
STATus Subsystem