Section
3: Functions and features Model DMM7510 7½ Digit Graphical Sampling Multimeter
Reference Manual
3-32 DMM7510-901-01 Rev. B / May 2015
Accessing the data in buffers
Using SCPI commands:
To access a buffer, include the buffer name in the respective command. For example, the following
commands:
Create a buffer named testData to store 100 readings
Set the instrument to make 5 readings for all measurement requests
Make the readings and store them in the buffer
Return five readings (including the measurement and relative time) from the user-defined buffer
named testData
TRAC:MAKE "testData", 100
SENS:COUN 5
TRAC:TRIG "testData"
TRAC:DATA? 1, 5, "testData", READ, REL
Using TSP commands:
A reading buffer is based on a Lua table. When you use TSP commands, the measurements
themselves are accessed by ordinary array notation. If rb is a reading buffer, the first measurement is
accessed as rb[1], the ninth measurement as rb[9], and so on. The additional information in the
table is accessed as additional members of the table.
To access a buffer, include the buffer name in the respective command. For example, the following
commands:
Create a buffer named testData to store 100 readings
Set the instrument to make 5 readings for all measurement requests
Make the readings and store them in the buffer
Return five readings (including the measurement and relative time) from the user-defined buffer
named testData
-- Create a buffer named testData to store 100 readings.
testData = buffer.make(100)
-- Set the instrument to make 5 readings and store them in the buffer.
trigger.model.load("SimpleLoop", 5, 0, testData)
-- Make the readings
trigger.model.initiate()
waitcomplete()
-- Read the 5 readings and print them including the measurement
-- and relative time for each reading.
printbuffer(1, 5, testData.readings, testData.relativetimestamps)

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB

Related product manuals