Model DMM7510
7½ Digit Graphical Sampling Multimeter Reference Manual Section 3:
Functions and features
DMM7510-901-01 Rev. B / May 2015 3-41
Storing configuration index 1
For example, use the following information to store configuration index 1 to MyMeasList.
You cannot mix digitize and measure functions in a configuration list that is used in a trigger model.
Using the front panel to configure the instrument:
Configure the instrument with the settings you want to store in the configuration index.
1. Press the HOME key.
2. Select the ACV function.
3. Select 10 V range.
4. Press the MENU key.
5. Under Measure, select Settings.
6. Set Detector Bandwidth to 3 Hz.
Using the front panel to store active settings to configuration index 1:
Store all active measure settings to MyMeasList as configuration index 1.
1. Press the MENU key.
2. Under Measure, select Config Lists. The MEASURE CONFIGURATION LISTS screen is
displayed.
3. Choose Select List. A menu of available configuration lists is displayed.
4. Select MyMeasList.
5. Select System to Index. This saves the active system settings to the configuration index. The
configuration index is displayed on the list.
6. Select Index Details to see some of the settings stored in configuration index 1.
Select HOME to return to the Home screen.
Storing configuration index 2
Refer to the instructions in Storing configuration index 1 (on page 3-41), and continue as follows to
configure the instrument for configuration index 2.
Using the front panel to configure the instrument:
Change the following instrument settings to configure the instrument with the settings you want to
save for configuration index 2:
Set the function to DC voltage
Set the range to 100 mV
Set the integration rate to 2 PLC
Using the front panel to store the active settings to configuration index 2:
1. Return to the configuration list menu.
2. Select MyMeasList.
3. Select System to Index. This saves the active system settings to the configuration index.
4. Select HOME to return to the Home screen.

Table of Contents

Other manuals for Keithley DMM7510

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley DMM7510 and is the answer not in the manual?

Keithley DMM7510 Specifications

General IconGeneral
Resolution7.5 digits
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range10 µA to 10 A
Resistance Range10 Ω to 1 GΩ
Sampling Rate1 MS/s
Memory27.5 million readings
Display5 inch capacitive touchscreen
AC Voltage Accuracy±0.06% of reading
AC Current Range10 A
ConnectivityUSB, LAN, GPIB