Chapter 5 Parameters Definition
132
When individual setting is different from global setting one or more of the following
appear:
F1 = Alarm Activity or F1 = Alarm Emission Rate
F1 = Cold Alarm Activity or F1 = Cold Alarm Emission Rate
F11 = High Alarm Activity or F11 = High Alarm Emission Rate
F3 = Alarm Test Source
F9 = Calibration Source
F7 = Measurement Process
F2 = Alarm Activity Dimension or F2 = Alarm Emission Rate Dimension
Line Items
Alarm Activity: See the F2 – (Zone) Settings screen on page 116.
Cold Alarm Activity: See the F2 – (Zone) Settings screen on page 116.
High Alarm Activity: See the F2 – (Zone) Settings screen on page 116.
Alarm Test Source (and related information) is used with Verification Test
otherwise it just documents the intended Alarm Test source. See F3 – Setup, F3 –
Detection Zones, F7 – Alarm Test Alarm Test Type setting. See the F2 – (Zone)
Settings screen on page 116.
The Source Activity, Date, and Dimension are automatically filled in based on the
Source selected.
Calibration Source (and related information) is used with Zone Calibration and
Optimize HV. See the F3 Calibration screen on page 121.
The Source Activity, Date, and Dimension are automatically filled in based on the
Source selected.
Measurement Process: When there is a difference with corresponding settings in
F3 Calibration then “F7 = Measurement Process ±” is available. See Bayesian
Supplementary Manual for more details.
Note this is actually applied to Calibration Process setting.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mirion Technologies GEM-5 and is the answer not in the manual?

Mirion Technologies GEM-5 Specifications

General IconGeneral
BrandMirion Technologies
ModelGEM-5
CategoryMonitor
LanguageEnglish

Related product manuals