Chapter 7 Retrieving Information
242
Figure 112 Service History Screen (page 2)
Function Keys
Ctrl+F4 = Clear History: At the discretion of your service support group, this
history may be kept for the life of the unit or may be reset periodically using the
Ctrl+F4 key, thus setting counters on both pages to zero and setting the “From...”
time to the current time. This is the only time the “From …” date/time is changed.
To prevent unauthorized or accidental use, a password must be entered to use this
function if Enable Access Level Control: is set to “YES” and the Level 2 or 3
Password is set in the F6 Access Level Settings screen (on page 37).
F3 Background History
The Background History for each detection zone (single detectors, single detector
sections if applicable and sum zones) is retained for the last 1080 background update
cycles in the D:\MonitorProgram\TEMPDATA\BKGND.DF file. These are the
values displayed on the Background History screen. A typical Background History
screen is shown in Figure 113.

Table of Contents