EasyManuals Logo

Mirion Technologies GEM-5 User Manual

Default Icon
443 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #315 background imageLoading...
Page #315 background image
Settings Files
Gem-5 User’s Manual 299
v_param.df
The v_param.df file contains the settings for the F6 Operation Voice & Screen.
These settings include voice choice and volume and whether set for single
language or multilingual voices and screens.
datalog.df
The datalog.df file stores the settings in the F7 Data Log screen. This includes
the settings for “*.csv” log files, such as frequency of logging, the maximum file
size and destination (file, printer, neither or both).
printout.df
The printout.df file contains the settings for the F8 – Printout screen that
determine whether various text files are created automatically and how many
days, they are retained.
netsetup.df
The netsetup.df file contains the settings for the network setup. These settings
are accessed through the F9 Data Transfer (Network) Setup screen.
instask.df, instbann.df, insttest.df, instclea.df, instcont.df, instfris.df,
instmess.df, instposi.df, instread.df, instserv.df, instturn.df
These *.df files contain the text for various Operation screens as set in the
F11 – Operation Instruction Menu screen.
alarmtst.df
The alarmtst.df file contains the settings for the Alarm Test setup. These settings
are accessed through the F7 Alarm Test Settings screen.
Test Menu
discrim.df
The discrim.df file contains the discriminator setting for each Detector Board.
These settings are accessed through the F8 Discriminator Plot screen.
plateau.df
The plateau.df file contains the detector high voltage setting for each detector.
High voltage settings are accessed through the F9 – Plateau Plot screen.
testamp.df
The testamp.df file contains the pulse amplitude and discriminator setting for the
amplifier pulse test. These settings are accessed through the F4 Amplifier
Pulse Test screen.
testrate.df
The testrate.df file contains the dwell setting for the rate meter test. This setting
is accessed through the F1 Rate Meter screen.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mirion Technologies GEM-5 and is the answer not in the manual?

Mirion Technologies GEM-5 Specifications

General IconGeneral
BrandMirion Technologies
ModelGEM-5
CategoryMonitor
LanguageEnglish

Related product manuals