Appendix B Operating Files and Directory Structure
298
timeout.df
The timeout.df file contains the settings for the F2 – Wait Timeouts screen.
z_value.df
The z_value.df file contains the detection zone parameter operation settings, as
set in the F1 – Zone Configuration screen. These settings include the Zone
Alarm and confidence (sigma) values that can be applied to all detection zones.
z_indivd.df
The z_indivd.df file contains the detector parameter operation settings, as set in
the F2 – (Zone) Settings screen. These settings include Zone Alarm; alarm
activity (release limit); dimensions and self shield factor; and background rate
alarm settings for every detection zone.
z_calib.df
The z_calib.df file contains parameter operation settings, as set in the
F3 – Calibration screen. These settings include calibration source data and
efficiency calibration results for every detection zone.
dc_param.df
The dc_param.df file contains the settings for the F5 – Group Settings screen.
These settings include the alarm activity (release limit) and calibration source
data for each group of detection zones.
z_global.df
The z_global file contains the settings for the Shift+F5 – Global Settings screen.
These settings include the alarm activity (release limit) and calibration source
data for each global detection group.
calibsrc.df
The calibsrc.df file contains the list of calibration sources entered into the
F9 – Edit Calibration Source List.
calsetup.df
The calsetup.df file contains the settings specified in the F3, F3, Shift+F3 -
Calibration Settings screen.
s_cause.df
The s_cause.df file contains the settings for the F4 – Out of Service Causes
screen, including the Recalibration Due Date/Time and Out-Of-Service
Instructions.
config.df
The config.df file contains the settings specified in the F5 – Configuration
screen.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mirion Technologies GEM-5 and is the answer not in the manual?

Mirion Technologies GEM-5 Specifications

General IconGeneral
BrandMirion Technologies
ModelGEM-5
CategoryMonitor
LanguageEnglish

Related product manuals